SETTING THE THERAPY INTENSITY

Section of the main LUX IV Instrument Panel
The THERAPY INTENSITY CONTROL knob is used to adjust the Lamp Transducer energy intensity for the patient’s size
and sensitivity. Small sensitive patients require less treatment intensity. Adjust the control so that the red Digital Volt Meter reads in the
correct range.
Children up to the age of 5, set the Treatment Intensity control knob so that the digital voltmeter reads between 3.00 and 6.00
volts.
Children from 5 to 9 years, between 6.00 and 9.00 volts.
Children from 9 years old, to adults, between 9.00 and 12.50 volts.
FOR ADVANCED USERS:
MANUALLY SETTING ANY DESIRED TREATMENT FREQUENCIES
 Section of the main LUX IV Instrument Panel
1. AUTOMATIC/MANUAL SWITCH, Press this switch to put the instrument into the manual mode. The switch will be
‘illuminated’. This function is for experienced or advanced users and permits the practitioner to set any frequency from 1.5 to 35,000
HZ.
2. FREQUENCY RANGE AND ADJUSTMENT KNOBS, Use these knobs to set a specific treatment frequency. For example to
set music note middle A which is 440 Hz, set the range switch to X100 and adjust the frequency control knob until the green Digital Frequency
Meter reads 440. Note: this meter is marked VIBRATIONAL RATE (HERTZ).
3. When setting frequencies below 25 Hz, press the SAMPLE 1S/10S switch so that the decimal point ‘illuminates’
on the green Digital Frequency Meter. This will allow greater low frequency setting accuracy by displaying the decimal point.